Version 1.04 of 8th May 2019
- [FW-550] Bug-Fix: Synchronized M2p cards not working with software trigger in Multiple Recording and ABA mode
- [FW-552] Bug-Fix: Cards with infinite recording/replay (e.g. FIFO modes) upon one single trigger might block Star-Hub from re-arming other connected cards
- [FW-555] Bug-Fix: Changes in Sync-Protokoll for replay created misaligned ABA A-channels
- [FW-561] Bug-Fix: Fast trigger rates can stop/lock starhub acquisitions
- [FW-563] Bug-Fix: Synchronization of cards with very different sample rates not working correctly
- [FW-564] Bug-Fix: Synchronization of cards with very different sample rates only works when fast card is trigger master